Reduction of Non-Actionable Alarms in Pulse Oximeters through Personalized Threshold Reparameterization
Files
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Access Restrictions
Abstract
The field of “smart alarms” has grown out of an influx of medical professionals demanding alarms be more representative of a need for direct medical intervention. In hospitals, many alarms indicate irrelevant physiological events, producing nuisance alarms. Herein we introduce two methods for non-actionable alarm reduction by learning from a patient's biometric data to intelligently set alarm parameters. Threshold variables are personalized for each patient, showing a rapid reduction in alarms of 68-88%. We focus on data from the device that accounts for most nuisance alarms in hospitals, the pulse oximeter. This framework presents a rapidly adoptable platform for reduction of alarms in medical devices generally.